CONFIGURING
THE METER
USING
SETPOINT 1 CONFIGURATION ("S1.CF")
Note: Refer to Table 6-1 for a summary list of menu configuration.
Setpoint 1 Configuration (S1.CF) is not active unless your meter
has dual relay output capabilities. The LED's will display whether the
("S1.CF") is active or not. You may use Setpoint 1 Configuration
("S1.CF") for the following:
* To Set setpoint's active band above or below your chosen value
* To Select whether the setpoint operation is latched or unlatched
Setting Setpoint 1's Active
Band
1. Press the MENU button until the meter shows "S1.CF".
2. Press the /TARE
button. The meter shows one of the following:
* "S.1=A" (Active above the deadband) (default)
* "S.1=B" (Active below the deadband)
3. Press the /MAX
button to toggle between available choices.
4. Press the /TARE
button to select if Setpoint 1 is latched or unlatched or press the MENU
button to store your selection.
Selecting if Setpoint 1 is
Latched or Unlatched
1. Press the "MENU" button until "S1.CF" displays,
then press the /TARE
button twice.
or Press the /TARE
button from "S.1".
The meter shows one of the following:
* "S.2=U" (setpoint 1 to be unlatched) (default)
* "S.2=L" (setpoint 1 to be latched)
2. Press the /MAX
button to toggle between available choices.
3.
Press the MENU button to store your selection(s). The meter momentarily
shows "STRD", followed by "S2.CF" (Setpoint 2 Configuration).
USING SETPOINT 2 CONFIGURATION
("S2.CF")
Note: Refer to Table 6-1 for a summary list of menu configuration.
Setpoint 2 Configuration ("S2.CF") is not active unless your meter
has dual relay output capabilities. The LED's will display whether the
("S2.CF") is active or not. You may use Setpoint 2 Configuration
("S2.CF") for the following:
* To set setpoint's
active band above or below your chosen value
* To select whether the setpoint operation is latched or unlatched
Setting
Setpoint 2's Active Band
1. Press the MENU button until the meter shows "S2.CF".
2.
Press the /TARE
button. The meter shows one of the following:
* "S.1=A" (Active above the setpoint) (default)
* "S.1=B" (Active below the setpoint)
3. Press the /MAX
button to toggle between available choices.
4. Press the /TARE
button to select if Setpoint 1 is latched or unlatched or press the MENU
button to store your selection and enter "S1.DB" (Setpoint 1 Deadband)
Selecting
if Setpoint 2 is Latched or Unlatched
1. Press the "MENU" button until "S2.CF" displays,
then press the /TARE
button twice.
or Press the /TARE
button from "S.1".
The meter shows one of the following:
* "S.2=U" (setpoint 1 to be unlatched) (default)
* "S.2=L" (setpoint 1 to be latched)
2. Press the /MAX
button to toggle between available choices.
3. Press the MENU button to store your selection(s). The
meter momentarily shows "STRD", followed by "S1.DB" (Setpoint
1 Deadband).
SETTING THE SETPOINT 1 DEADBAND
("S1.DB")
Note: Refer to Table 6-1 for a summary list of menu configuration.
Setpoint 1 Deadband ("S1.DB") is not active unless your meter
has dual relay output capabilities. The LED's will display whether the
("S1.DB") is active or not. The Setpoint 1 default deadband is
0003. To change the deadband (hysteresis) of setpoint 1, follow these
steps:
1. Press the MENU button until the meter shows "S1.DB".
2. Press the /TARE
button. The meter shows the last previously stored 4-digit number (0000
through 9999) with flashing 4th digit.
3. Press the /MAX
button to change the value of the flashing digit. If you continue to press
the /MAX button,
the flashing digit's value continues to change.
4. Press the /TARE
button to scroll to the next digit.
5. Press the MENU button to store your selection. The meter
momentarily shows "STRD", followed by "S2.DB" (Setpoint
2 Deadband).
SETTING
THE SETPOINT 2 DEADBAND ("S2.DB")
Note: Refer
to Table 6-1 for a summary list of menu configuration.
Setpoint 2 Deadband
("S2.DB") is not active unless your meter has dual relay output
capabilities. The LED's will display whether the (S2.DB) is active
or not. The Setpoint 2 default deadband is 0003. To change the deadband
(hysteresis) of setpoint 2, follow these steps:
1. Press the MENU button until the meter shows "S2.DB".
2. Press the /TARE
button. The meter shows the last previously stored 4-digit number (0000
through 9999) with flashing 4th digit.
3. Press the /MAX
button to change the value of the flashing digit. If you continue to press
the /MAX button,
the flashing digit's value continues to change.
4. Press the /TARE
button to scroll horizontally to the next digit.
5. Press the MENU button to store your selection. The meter
momentarily shows "STRD", followed by "OT.CF" (Output Configuration)
if you have analog output capabilities.
Alarm
Example
Figure 4-1
To reset
latched alarms you must:
1. Input a signal "OUT" of the alarm
zone
2. Then press SETPTS and then, RESET button |
|